Interchangeability of cardiac output measurements between non-invasive photoplethysmography and bolus thermodilution: A systematic review and individual patient data meta-analysis

Résumé

Background: Continuous non-invasive cardiac output devices using digital photoplethysmography (PPG) are widely available for bedside use, but their interchangeability with reference methods has not yet been evaluated in a systematic review and patient data meta-analysis. Methods: A systematic review and meta-analysis of studies comparing non-invasive cardiac output monitoring using PPG with the invasive bolus thermodilution method was performed. With ethical approval, all published studies from the PUBMED, Embase, Scopus, Web of Science, and Google Scholar databases from January 1, 2010 to January 1, 2018 were included. From these analysed studies, individual patient data were interpreted using the interchangeability methods for both absolute values and changes in cardiac output measurements. Results: Ten studies comparing PPG and bolus thermodilution in the operating room and intensive care settings were included. The interchangeability rate (95% CI) was 37% (24-48) (n = 1350 pairs of measurements). The interchangeability rate was poorer with the CHAP device (CNSystems, Graz, Austria) [18% (17-20)] than with the Clearsight (Edwards Lifesciences, Irvine, CA) device [33% (31-34), P < 0.0001], for patients receiving norepinephrine [19% (17-20) vs. 33% (32-34), P < 0.0001], and for patients with low mean arterial pressure (< 65 mmHg) [26% (23-29) vs. 30% (29-31), P < 0.0001]. Among the 1009 comparisons of the changes in cardiac output between both methods, 561 (56%) were interpretable with a trend interchangeability rate at 24% (12-36). Conclusions: Cardiac output measurements using PPG were not interchangeable with bolus thermodilution in regard to both absolute values and changes in cardiac output measurements, and should be used with caution in clinical practice. (C) 2019 The Authors. Published by Elsevier Masson SAS on behalf of Societe francaise danesthesie et de reanimation (Sfar). This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

Dates et versions

hal-03636190 , version 1 (09-04-2022)

Identifiants

Citer

Marc-Olivier Fischer, Alexandre Joosten, Olivier Desebbe, Mariam Boutros, Stephane Debroczi, et al.. Interchangeability of cardiac output measurements between non-invasive photoplethysmography and bolus thermodilution: A systematic review and individual patient data meta-analysis. Anaesthesia Critical Care & Pain Medicine, 2020, 39 (1), pp.75-85. ⟨10.1016/j.accpm.2019.05.007⟩. ⟨hal-03636190⟩
55 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More